Pre-syncope during exercise in a patient with a permanent pacemaker and hypertrophic obstructive cardiomyopathy

 

Hayley Langridge – Advanced Specialist Cardiac Physiologist

Manchester University NHS Trust.

hayley.langridge@mft.nhs.uk

 

Background

A 71-year-old male with a history of hypertrophic obstructive cardiomyopathy and dual chamber pacemaker implantation for symptomatic 2:1 AV block, presented to A&E with dizzy spells and near syncope during exercise. A pacemaker check was performed in A&E with the device function deemed satisfactory. The patient was later referred for an exercise test to rule out exertion induced arrhythmias and assess pacemaker behaviour during exercise.  Figure 1 shows the EGM recorded during the exercise test at approximately 10 minutes of exercise which was associated with symptoms of pre-syncope and lethargy.  The device settings are displayed below.

 

Mode: DDDR

LR/UR: 60/130 bpm.

Paced/Sensed AV delay: 180/140 ms

PVARP: Auto

PVAB: 150 ms Partial +

Mode switch: On                   

Sensitivity thresholds: Atrial 0.3 mV, Ventricular 0.9 mV

 

Figure 1

EGM recorded 10 minutes into exercise at the time of symptoms. Channel 1 shows the Can to RV ring EGM with event marker channels. Channel 2 shows a poor quality atrial EGM and Channel 3 shows the RV tip to RV ring EGM.

 

Question

From the options below what programming change would you make?

□ No change normal device function

□ Increase max tracking rate

□ Shorten PVARP

□ Reduce ventricular sensitivity

□ Shorten PVAB

Answer 

Click here