September 2020

A Remote Interrogation requiring further Investigation

Authors:
Dr Farhana Ara – Specialist ST7 Registrar in Cardiology: Devices and Heart Failure
Sandra Silva – Highly Specialised Cardiac Physiologist

Disclosure: The author has no conflict of interests to declare.

Case presentation

  • 81M with previous atrial fibrillation underwent a VVI PPM (Medtronic Vivatron G20A2 SR) implantation for complete heart block. At the post-implant check and six-week in-house device interrogation, the RV threshold was 0.5V@0.4ms.
  • The first remote transmission (five months post-implant) showed a gradual increase in threshold from the day of implant.
  • The RV threshold was 2.5V@0.4ms and the output was automatically set to 5V@1.0ms.

Device settings
Mode: VVIR
Lower rate: 60ppm
Upper sensor rate: 130ppm
Pace/Sense polarity: Bipolar/bipolar
Output management: Adaptive

Figure 1: RV lead threshold

QUESTION 1

What would be next step in management?

  • Continuing monitoring

  • CXR

  • In-house device interrogation

  • List for lead revision

Case Continued
The patient was listed for a lead revision. Fluoroscopy of the lead and device interrogation on the morning prior to procedure showed no lead noise or displacement, with a stable lead impedance.

Figure 2: RV lead impedance trends

A manual threshold test was then performed which showed loss of capture at 0.25V@0.4ms, giving a manual threshold of 0.5V@0.4ms.

Figure 3: Manual threshold test

An automated threshold test was also performed.

Figure 4: Automated threshold test

QUESTION 2

What is the next step in management?

(ensure you read answer to question 1 first)

  • Add an atrial lead

  • List for lead revision

  • Monitor

  • Switch off autocapture

Answer
Autocapture adaptive pacing was switched off. There was a consistently normal RV threshold (0.5V@0.4ms). Inappropriate capture management was noted, and the impression was of inappropriate evoked response due to hyperpolarization.
Lead revision was cancelled, and the patient was well at the 3-month follow-up appointment.

Discussion

  • Ventricular capture management is an automatic pacing threshold adjustment algorithm that automatically measures pacing threshold through detection of the evoked response after a pacing stimulus.
  • This algorithm then automatically updates the pacing output to optimal value. This function avoids unnecessary high output pacing and is considered to prolong battery longevity.
  • Pacing threshold search is initiated according to the schedule and frequency programmed by the clinician in pacemakers (nominally set to ‘day at rest’).
  • After implant detection complete: the device initiates the first pacing threshold search 12 hours after.
  • If a search cannot be completed, the device retries after 30 minutes.
  • A ‘High Threshold Condition’ warning is issued if the amplitude threshold is greater than 2.5V. The pacemaker responds by adapting to an amplitude of 5.0V and pulse width of 1.0ms. It will not give you an exact RV threshold measurement (Figure 1).

 

Figure 5: Ventricular capture management

  •  Figure 5 shows ventricular capture management with a pacing stimulus or ‘test pulse’ being delivered to the right ventricular. When there is no evoked response detected during the pacing search within a timed interval, back-up pulse is delivered, followed by 3 further pacing pulses or ‘support events’.
  • Capture verification efficiency relies on measurement of the evoked response and works by triggering a blanking window (closed) followed by an open window.
  • If the device detects an ER within this second window, capture is confirmed (Figure 6). Otherwise it will deliver a 5V back-up pulse to ensure capture (Figure 7).

Figure 6: Capture verification

 

Figure 7: Capture verification with back-up pulse

 

Figure 8: Ventricular capture management in our patient

  • Figure 8 shows that the device is allowing 3 support cycles as programmed rate and output followed by a faster test pace.  The test pace automatically has a backup pace at programmed voltage at 1.0ms that occurs 110msec after the test pace.  These backup pulses are not seen on this strip.
  • This cycle repeats looking for the 2/3 capture before moving on or 3/3 capture to confirm measurement.
  • Looking at the timing of the rhythm then it seems that all test paces capture as the ventricular event does not shift 110msec later.  However, the evoked response signal is not consistently detected by the device so that it sometimes labels this as LOC.
  • Evoked response detection and limitations: In rare instances, the pacemaker may not detect the electronic waveform created by the contracting myocardium immediately following a pacing pulse.
  • High thresholds may be appropriate and due to lead dislodgement, or incomplete connection of the lead in the header block.
  • Acute effects of lead maturation can also cause failure to detect the evoked response. Hyperpolarization of the EGM leading to signal saturation or changes in lead tip and muscle interface (i.e. tissue fibrosis) may be responsible for this.
  • In our patient, the phenomenon was seen following the lead maturation period (typically within 4-12weeks) when fibrosis between tip to tissue interface is expected. The inflammatory reaction and the resulting fibrous tissue that occurs after implantation may act as an insulating shield around the electrode, effectively raising the threshold for stimulation.
  • Remote transmission has increased during the COVID 19 pandemic. This case highlights the importance of inpatient/face-to-face review and interpretation of automated diagnostics. 

With acknowledgements to Dr Greg Mellor, Consultant Cardiologist and Electrophysiologist, Royal Papworth NHS Trust


References
  1. Antretter H, Bonatti J, Cottogni M, et al. Eine neue Herzschrittmacher-Stimulationstechnologie (Autocapture) ermöglicht bei unveränderter Lebensdauer die Reduktion der Generatorgrösse auf die Hälfte [A new cardiac pacemaker stimulation technology (autocapture) allows, with unchanged life expectancy, reduction of generator size by half]. Acta Med Austriaca. 1996;23(5):159-164.
  1. https://www.medtronicacademy.com/features/ventricular-capture-management-feature
  1. Kishihara, Jun & Niwano, Shinichi & Fukaya, Hidehira & Nishinarita, Ryo & Horiguchi, Ai & Nakamura, Hironori & Igarashi, Tazuru & Ishizue, Naruya & Fujiishi, Tamami & Yoshizawa, Tomoharu & Oikawa, Jun & Satoh, Akira & Murakami, Masami & Ako, Junya. (2017). Pacing failure caused by automatic pacing threshold adjustment system. Journal of Arrhythmia. 33. 10.1016/j.joa.2017.05.005.
  1. Hayes D, Pacing system malfunction: Evaluation and management 2019 https://somepomed.org/articulos/contents/mobipreview.htm?33/20/34113/contributors
  1. HRS/EHRA/APHRS/LAHRS/ACC/AHA worldwide practice update for telehealth and arrhythmia monitoring during and after a pandemic (2020) https://doi.org/10.1016/j.hrthm.2020.06.010
  1. Murgatroyd FD, Helming E, Lemke B et al. Manual vs. automatic capture management in implantable cardioverter defibrillators and cardiac resynchronization therapy defibrillators Europace (2010) 12, 811–816 doi:10.1093/europace/euq053