January 2020

Unscheduled remote transmission that unravelled a lead issue

Phil Matthews – Highly Specialised Cardiac Physiologist
The Rotherham NHS Foundation Trust
phillip.matthews@nhs.net

Background
A 71 year old female patient was implanted with a St Jude Medical Quadra Assura MP 3371-40QC CRT-D device for heart failure symptoms (SOBOE, Reduced Exercise Tolerance – classified NYHA III) in the context of Dilated Cardiomyopathy and Severe LVSD (EF 10%). ECG performed in the physiologist led Pre-Assessment Clinic showed Sinus Rhythm 71bpm with Left Bundle Branch Block (QRS duration 160ms); 6 minute walk test was performed (distance achieved of 185 metres) and a Minnesota QoL questionnaire completed (scored 56/100). The implant procedure was successful with a St Jude Medical Quartet 1458Q lead placed in a Posterolateral branch of the Coronary Sinus. Post-implant checks were satisfactory and QRS duration had been reduced to 131ms through the use of SJM’s CRT Optimisation toolkit. The device settings programmed at implant are displayed below:-

Brady settings

Mode:                  DDD
Base Rate:          50bpm
Max Track:        120bpm
Paced AVD:       225ms
Sensed AVD:     200ms
SyncAV Delta:  -30ms
LV-RV Offset:   40ms

Tachy settings

VT-1 Zone             160bpm 375ms             40 Intervals           Monitor Zone
VT-2 Zone            200bpm 300ms            30 Intervals           ATP x 3; HV x 4
VF Zone                240bpm 250ms             8 Intervals             ATP during charging; HV x 6

Morphology On
Onset On
Stability On

If All Indicate VT, deliver therapy

18 days later we received an alert via the Merlin.net Patient Care Network of BiV pacing percentage less than limit (<90%) and RV CapConfirm in high output mode. The real-time EGM of the transmission was checked to establish the potential cause of RV CapConfirm being in high output mode.

Figure 1
Real-time EGM of the transmission received 18 days post clinic visit

QUESTION

Which answer best explains the above EGM recording?

  • T-Wave Oversensing

  • Pacemaker Crosstalk

  • Ventricular Lead Displacement
  • Ventricular Bigeminy

Answer
Ventricular Lead Displacement

Explanation
The real-time EGM shows ventricular lead displacement.
The EGM shows a single sensed event on the atrial channel while the ventricular channel shows 2 sensed events. Every VS is followed by a subsequent sensed episode 180-188ms later that is represented with a -, indicating that the event is binned and does not count towards detecting a VT or VF episode. The Sensed AV interval is very short – varying between 27-35ms – and is non-physiological in duration and being indicative of a far field signal. Therefore, the ventricular lead is likely to have displaced and has migrated in-between the tricuspid valve allowing it to sense both atrial and ventricular activity.

This is different from pacemaker crosstalk which is when an atrial paced event is sensed on the ventricular channel, causing inhibition of ventricular pacing. This can be rectified by the use of ventricular safety pacing (VSP) which detects for the presence of crosstalk by monitoring for non-physiologic ventricular sensed events (e.g. an ApVs interval of less than 110ms on Medtronic devices) and responds by pacing the ventricle.

Action
The patient was contacted and was asked to attend the Cardiac Devices Suite as a matter of urgency, however she was unable to attend till the following day. The next day, the following episode occurred as the patient was en-route to the hospital.

Figure 2
Stored EGM of an episode as the patient was en-route to the hospital

The above episode – amongst 5 other similar episodes – shows inappropriate therapy that has occurred due to the lead displacement. In the first EGM strip, the sinus node rate increases therefore this decreases the interval between the true ventricular activation and the “far-field” atrial activation on the ventricular lead. St Jude Medical high energy devices detect tachycardia episodes by comparing the current R-R interval to the average R-R interval, calculated by (sum of current interval + previous 3 intervals) / 4. If either the current or the average intervals is measured outside of the tachy detections zones (in this case, >375ms), that interval is discarded. We can see that every V-A interval on the RV channel is greater than 375ms, therefore these intervals are discarded as the current interval falls out of the tachy detection zones. However the A-V intervals (measured approximately 150-180ms) are binned in the VF zone as both the current and the average interval falls within the tachy detection zones.

At the ‘star’ marker, we see that while the atrial activation has been sensed on the ventricular lead, ventricular activation was not sensed. This results in the average interval rate for the next 4 events to be measured outside the tachy detection zones, therefore the intrinsic V activity will not be binned into either the VT-2 or VF treatment zones. From the second ‘arrow’ marker onwards, the average interval is now fast enough to fall into the detection zones, therefore binning of these events in the VF zone continues with the 8th event binned at the end of the 1st strip.

In the 2nd EGM strip, at the ‘arrow’ marker both the current and the average intervals are within the re-detection zones and binning in the VF zone commences again after ATP during charging. The sequence of events previously described continues onto the 3rd EGM strip until the 6th episode is binned in the VF zone is reached (denoted by the orange arrow), meaning VF has been redetected by the device and upon occurrence of the 7th event, HV therapy is delivered.

Upon arrival into A&E, tachy therapies were disabled and the patient was referred for an urgent CXR displayed in figure 3

Figure 3
X-ray performed in A&E on the left. Post implant X-ray on the right.

CXR on the left shows ventricular lead displacement in comparison to the post implant CXR. The cause of the ventricular lead displacement is due to Twiddler’s Syndrome which can be seen by the presentation of the leads in the device pocket. The next day the patient underwent urgent right ventricular lead replacement and has since then remained event free.

Twiddler’s Syndrome
Twiddler’s Syndrome occurs due to intentional or unintentional manipulation of the device pulse generator within the skin pocket by the patient. This manipulation causes coiling of the lead in the pocket and its dislodgement from its implanted position, resulting in failure of ventricular pacing and potential sensing abnormalities. The majority of cases are diagnosed within the first year of implant and has an estimated frequency of around 0.07-7% [1]. Manifestation of Twiddler’s Syndrome can vary and dependent on the displacement of the lead, can result in unintentional stimulation of the ipsilateral phrenic nerves or the brachial plexus [2]. Risk factors for the syndrome include obesity, elderly age; female gender, impaired cognition and a large sized device pocket relative to the size of the device generator [3].

Often patients will report a sudden return of their pre-implant symptoms (in cases where devices were implanted for symptomatic bradycardia and AV block), however in patients where devices were implanted for rarely occurring bradyarrhythmias or for prophylactic reasons, this may not be the case. Abnormal lead measurements such as a reduced sensing amplitude and high threshold or non-capture will be present at a device follow-up and a chest x-ray – performed to evaluate lead positioning – will show evidence of the lead coiling in the device pocket.

Treatment for cases of Twiddler’s Syndrome involves uncoiling and repositioning or replacement of the affected lead with repositioning of the pulse generator. Suture fixation of the device generator and the leads with a smaller device pocket without redundant space can prevent the occurrence of Twiddler’s Syndrome, however too small of a device pocket can be a predisposition to device erosion.
References:

[1] Salahuddin M, Cader FA, Nasrin S, Chowdhury MZ. The pacemaker-twiddler’s syndrome: an infrequent cause of pacemaker failure. BMC Res Notes. 2016;9:32.

[2] DeMarco DC, Xuereb RG. ‘Twiddling’ of the pacemaker resulting in lead dislodgement. Malta Med J. 2009;21(3):38–41

[3] Castillo R, Cavusoglu E: Twiddler’s Syndrome: An Interesting Cause of Pacemaker Failure. Cardiology 2006;105:119-121